Veris Technologies VIS_NIR Spectrophotometer - Operating Instructions User Manual

Page 74

Advertising
background image

Pub# OM17-NIR Shank

5-21

Procedure #7 Replacing halogen lamp

1. Remove the shank from the implement following procedure on page 4-11

2. Remove the top plate on the shank by unscrewing four screws

Figure 22

USE CAUTION WHEN HANDLING TOP COVER! TOO MUCH MOVEMENT
CAN FLEX AND BREAK THE FIBER OPTIC CABLE.

3. Using a 3/32 hex wrench loosen the set screw holding in the lamp housing.

Figure 23

Advertising