Pub# OM17-NIR Shank

5-14

o Control Port
 Bits per second: 9600
 Data bits: 8
 Parity: None
 Stop bits: 1
 Flow control: None

o Spectrometer Port
 Bits per second: 9600
 Data bits: 8
 Parity: None
 Stop bits: 1
 Flow control: none

Port Definitions

o Temperature Port

– Reads in soil temperature.

 Example of temperature port connected in HyperTerminal


o GPS Port

– Reads in GPS coordinates.

Example of GPS port connected in HyperTerminal.

o Control Port

– Reads in EC, shutter position, Auxiliary case temperature

and humidity, and shutter steps. After opening this port if no data is
present type S for shank mode or P for probe mode. The letter after E
represents the shutter state. In the case below the shutter is in neutral or
rest. If the case reads R the shutter is in reference or if there is a D the
shutter is in dark. To manually move the shutter type N for neutral, then R
for reference, then D for dark. The shutter can only be moved in this
order.