VEGA PLICSCOM User Manual

Page 19

Advertising
background image

19

5 Setup - Sensors up to software 3.99

PLICSCOM •

36433-EN-130228

The sensor is already set to the ordered national language. In this

menu item you can change the language. The following languages are

available, e.g. in software version 3.50:

Deutsch

English

Français

Espanõl

Pycckuu

Japanese

Italiano

Netherlands

Japanese

Chinese

Language

German

With this function

Load parameter adjustment data from the sensor into the display

and adjustment module

Write parameter adjustment data from the display and adjustment

module into the sensor

The data are permanently saved in an EEPROM memory in the dis-

play and adjustment module and remain there even in case of power

failure. From there, they can be written into one or several sensors or

kept as backup for a possible sensor exchange.
The type and the volume of the copied data depend on the respective

sensor.
Information:

Before data are written into the sensor, a check is carried out to

determine whether the data fit the sensor. If the data do not fit, a fault

signal is triggered or the function is blocked. When writing data into

the sensor, you will see which instrument type the data originate from

and which TAG-no. this sensor had.
The following items are checked:

Software version

WHG approval

SIL activated

Measuring principle

Radar C-band/K-band

Radar measuring range < 30 m or > 30 m

Signal output

Pressure measuring range

Copy sensor data

Copy sensor data?

Service/Language

Service/Copy sensor data

Advertising