VEGA PLICSCOM User Manual
Page 34
34
6 Setup - radar sensors from software 4.0.0
PLICSCOM •
36433-EN-130228
are carried out. In this menu item, the value fir the function block is
selected:
•
PV (Primary Value):
– Linearised percentage value
•
SV1 (Secondary Value 1):
– Percent with radar, guided microwave and ultrasonic sensors
– Pressure or height with pressure transmitters
•
SV2 (Secondary Value 2):
– Distance with radar, guided microwave and ultrasonic sensors
– Percent with pressure transmitters
•
Height
Additional scalings (Out-Scale) are carried out within the function
block. In this menu item, the scaling unit is selected. For a better
overview, the units are combined in groups:
Additional scalings (Out-Scale) are carried out within the function
block. In this menu item, the unit of the scaling is entered.
To damp process-dependent measured value fluctuations, set an
integration time of 0 … 999 s in this menu item.
Depending on the sensor type, the factory setting is 0 s or 1 s.
Setup/AI FB1 Scaling unit
Setup/AI FB1 Scaling
Setup/AI FB1 damping