Fault finding

Shuttle

®

indicates

system error

constantly

Shuttle

®

Setting

Is the setting of the active measuring range (Shift + B)

correct? The active measuring range must not be set

lower than the max. possible level.

Level read-

out is not

changing

Ultrasonic

sensor

Sensor

Is the sensor mounted ABSOLUTELY VERTICAL?

mounting

It is extremely important that the sensor is firmly

mounted in a vertical position. See the section

'Mechanical mounting of sensor'.

Installation

Are there for example bit, fatty accumulations or other

site

objects disturbing the ultrasonic signal?

Setting

Has the learning function been activated? If the

learning function has not been activated, Shuttle

®

may

lock on a false echo.

Level read-

out is wrong

Ultrasonic

sensor

Sensor

Is the sensor mounted ABSOLUTELY VERTICAL?

mounting

It is extremely important that the sensor is firmly

mounted in a vertical position. See the section

'Mechanical mounting of sensor'.

Cable

Is the sensor cable extended with a not-approved

cable type and/or extended beyond 100 m?

Installation

Does the ultrasonic sensor have the same temperature

as the ambient air? Deviations will produce measuring

errors!

Shuttle

®

Setting

Shuttle

®

level readout may need adjustment.

Level

read-out is

unstable

Ultrasonic

sensor

Sensor

Is the sensor mounted ABSOLUTELY VERTICAL?

mounting

It is extremely important that the sensor is firmly

mounted in a vertical position. See the section

'Mechanical mounting of sensor'.

Is the sensor mounted firmly? The sensor should be

mounted on a suitable bracket.

Installation

Turbulence on the surface.

site

Objects on the surface that disturb the measurement.

Strong winds can bend off the echo, so it misses the

ultrasonic sensor.

Shuttle

®

Setting

Response time (Shift + C) is set too low.