Mounting

Electrical mounting
Level transmitter

The Shuttle

®

must not be connected to

the power supply before the ultrasonic

sensor is mounted and connected cor-

rectly.

When the cover has been removed, the
green plastic film with the menu symbols is
tipped up to gain access to the terminals.

Shuttle

®

can be supplied with 10 - 30 V DC

on terminal 10 and 12 or with 230 / 115 V
AC on terminal L and N.

Current regulations for conductor and

fuse dimensions should always be ob-

served.

Always confirm that the Shuttle

®

voltage

rating match the present voltage.
If Shuttle

®

is delivered for 115 V AC supply,

it will be indicated with a label below the
leftmost terminal block as shown here:

Shuttle

®

is intended

Shuttle

®

is intended

for 230 V AC.

for 115 V AC.

Terminal: Designation:
1 - 5

Ultrasonic sensor

6 - 7

Relay output 1

(Max. 50 V, 1 A resistive load)

8 - 9

Relay output 2

(Max. 50 V, 1 A resistive load)

10 and 11 4-20 mA output

(Max. 500

Ω load)

10 and 12 10 - 30 V DC supply

L

230 / 115 V AC live

N

230 / 115 V AC neutral

Mount the wires according to the terminal
numbers on the reverse side of the green
plastic film: