Controls and Indicators

2-2

LIFEPAK 1000 Defibrillator Operating Instructions

CONTROLS AND INDICATORS

This section introduces you to the controls and indicators on the LIFEPAK 1000 defibrillator.

Figure 2-1

Controls and Indicator

Table 2-1

Controls and Indicators

Feature

Description

1 Readiness

display

The

readiness display alerts you to the defibrillator’s readiness

status.

Three symbols (

,

,

) allow you to determine whether the

defibrillator is ready for use or needs attention.

The following defines what each symbol represents and
when/where each appears.

The wrench indicator appears on the readiness display when a
condition exists that prevents or could prevent normal
defibrillator operation.

The OK symbol indicates that the defibrillator is ready for use.

This symbol is visible only when the defibrillator is off.

The battery symbol appears on the readiness display when the
defibrillator is off. When one bar is visible in the symbol, the
battery is low. If the symbol is blank, the battery is extremely
low and the OK symbol will not appear when the defibrillator is
off.

2

Speaker

Provides audio voice prompts and tones.