3-10

LIFEPAK 1000 Defibrillator Operating Instructions

How to Use the LIFEPAK 1000 Defibrillator

Troubleshooting Tips for ECG Monitoring

If problems occur while monitoring the ECG, check this list of observations for troubleshooting
assistance.

Table 3-2

Troubleshooting Tips for ECG Monitoring

Observation

Possible Cause

What to Do

Screen blank and

ON

LED

lit.

Screen not functioning
properly.

• Contact authorized service

personnel.

• AED and therapy functions may

still operate. If needed for
therapy, continue to use device
to treat patient.

CONNECT ECG LEADS

message appears

One or more ECG electrodes
are disconnected.

• Confirm ECG electrode

connections.

Poor electrode-to-skin
contact.

• Reposition cable and/or lead

wires to prevent electrodes from
pulling away from patient.

• Clean, shave, and dry the

patient’s skin as recommended
on

page 3-3

.

• Replace electrodes.
• Change cable.

Broken ECG cable lead wire.

• Check ECG cable continuity. If

lead wire is broken, replace ECG
cable.