Physio-Control LIFEPAK 1000 User Manual

Page 33

Advertising
background image

LIFEPAK 1000 Defibrillator Operating Instructions

3-11

©2006-2012 Physio-Control, Inc.

How to Use the LIFEPAK 1000 Defibrillator

Poor ECG signal quality.

Poor electrode-to-skin
contact.

• Reposition cable and/or lead

wires to prevent electrodes from
pulling away from patient.
Secure cable clasp to patient’s
clothing.

• Clean, shave, and dry the

patient’s skin as recommended
on

page 3-3

.

• Replace electrode(s).

Outdated, corroded, or
dried-out electrodes.

• Check date codes on electrode

packages.

• Use only silver/silver chloride

electrodes with Use By dates
that have not passed.

• Leave electrodes in sealed

packet until time of use.

Loose connection.

• Check/reconnect cable

connections.

Damaged cable or
connector/lead wire.

• Inspect ECG and therapy

cables.

• Replace if damaged.
• Check cable with simulator and

replace if malfunction observed.

Noise because of radio
frequency interference (RFI).

• Check for equipment causing

RFI (such as a radio transmitter)
and relocate or turn off
equipment power.

Baseline wander
(low frequency/high
amplitude artifact).

Inadequate skin preparation.
Poor electrode-to-skin
contact.

• Clean, shave, and dry the

patient’s skin as recommended
on

page 3-3

.

• Replace electrodes.

Fine baseline artifact
(high frequency/low
amplitude).

Inadequate skin preparation.
Isometric muscle tension in
arms or legs.

• Clean, shave, and dry the

patient’s skin as recommended
on

page 3-3

.

• Replace electrodes.
• Confirm that limbs are resting on

a supportive surface.

• Check electrodes for proper

adhesion.

Table 3-2

Troubleshooting Tips for ECG Monitoring (Continued)

Observation

Possible Cause

What to Do

Advertising