Auto tests, Inspection, Auto tests -3 – Physio-Control LIFEPAK 1000 User Manual

Page 41: Inspection -3

Advertising
background image

Caring for the LIFEPAK 1000

LIFEPAK 1000 Defibrillator Operating Instructions

5-3

©2006-2012 Physio-Control, Inc.

Auto Tests

The defibrillator performs automatic self-tests daily and monthly at 0300 (3:00 a.m.) if not in use.
During the automatic self-test, the defibrillator turns itself on (ON/OFF LED illuminates) briefly
and completes the following tasks:

• Performs a self-test

• Stores the self-test results in the Test Log

• Turns itself off

If the defibrillator detects a problem during an auto test that requires service, it displays the
service symbol. If the service symbol is visible, you should attempt to use the defibrillator, if
needed, for a cardiac emergency. However, you should contact authorized service personnel to
correct the problem as soon as possible. The service symbol will remain visible until the
problem is corrected.

The automatic self-test is not performed if the defibrillator is already turned on at 0300 or if the
battery is not installed. If the defibrillator is turned on while a self-test is in progress, the test is
halted; the defibrillator will turn on normally.

INSPECTION

Routinely inspect all devices, accessories, and cables by following the instructions in Table 5-2.

Table 5-2

LIFEPAK 1000 Defibrillator Inspection

Instruction

Inspect For

Recommended Corrective Action

Examine the
defibrillator case,
connector, battery well,
battery pins, and
accessories.

Foreign substances.

Clean the device as described in

Table 5-3

.

Damage or cracks.

Contact authorized service personnel
to troubleshoot.

Battery pins bent or
discolored.

Contact authorized service personnel.

Expired batteries or
defibrillation electrodes.

Replace.

Observe readiness
display

OK

symbol

None needed.

Low or replace battery
indication displayed

Replace battery immediately.

Service symbol displayed

Contact authorized service personnel.

Examine accessory
cables.

Foreign substances.

Clean the cables as described in

Table 5-3

.

Inspect for cracks, damage,
extreme wear, broken or bent
connectors and pins.

Replace damaged or broken parts.

Confirm that connectors
engage securely.

Replace damaged or broken parts.

Advertising