About cprmax technology, Aed operation with cprmax technology, Initial cpr – Physio-Control LIFEPAK 1000 User Manual

Page 65

Advertising
background image

cprMAX™ Technology

LIFEPAK 1000 Defibrillator Operating Instructions

C-1

©2006-2012 Physio-Control, Inc.

ABOUT cprMAX TECHNOLOGY

Physio-Control cprMAX technology is designed to allow resuscitation protocols to maximize the
quantity of CPR administered during treatment with an AED, consistent with the 2005 American
Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency
Cardiovascular Care

1

(AHA Guidelines) and the European Resuscitation Council Guidelines for

Resuscitation 2005

2

(ERC Guidelines).

Setup options should be changed only under the direction of a physician knowledgeable in
cardiopulmonary resuscitation who is familiar with the literature in this area.

The cprMAX technology includes the following setup options:

INITIAL CPR

. Prompts the user to perform an initial period of CPR. Applies only to

immediately after turning on the AED or after the first analysis.

PRESHOCK CPR T

ime. Prompts for CPR after a shockable ECG rhythm is detected, before

the shock is delivered. If

INITIAL CPR

is set to

OFF

, then

PRESHOCK CPR

applies to all shock

advised decisions (including the first analysis).

CPR TIME 1

and

2

. CPR time periods after shocks or no shock advised decisions

respectively.

STACKED SHOCKS

. Eliminates the analysis after each shock and inserts prompting for CPR

after each shock; when set to

OFF

, eliminates the three-stack shock.

PULSE CHECK

. Indicates when, if ever, the device is to prompt for pulse checks.

AED protocols are aligned with the AHA and ERC Guidelines when the setup options are set as
follows:

• Initial CPR:

OFF

• PreShock CPR Time:

OFF

• CPR Times 1 and 2:

120 SECONDS

• Stacked Shocks:

OFF

• Pulse Check:

NEVER

The above options are the factory default settings for cprMAX technology. Your medical director
protocols should determine whether or not to change the options and should ensure that you
receive training.

AED OPERATION WITH cprMAX TECHNOLOGY

The following paragraphs describe AED operation with cprMAX technology setup options.

Initial CPR

The

INITIAL CPR

option prompts the user to perform an initial period of CPR. The choices are:

OFF

,

ANALYZE FIRST

and

CPR FIRST

. The factory default is

OFF

.

• The

OFF

setting has no prompting for an initial CPR period.

• The

ANALYZE FIRST

setting prompts for analysis and then CPR. If the analysis determines

1

2005 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular

Care. Circulation 2005;112 (Supplement IV).

2

European Resuscitation Council Guidelines for Resuscitation 2005. J. Resuscitation 2005; 67 (Supplement 1).

Advertising