Specifications

A-2

LIFEPAK 1000 Defibrillator Operating Instructions

Waveshape and Measured Parameters:

Patient
Impedance (

) I1 (A)

I2 (A)

I3 (A)

I4 (A)

T1 (ms)

T2 (ms)

25

50.3

20.1

19.7

10.7

5.9

3.9

50

28.2

14.6

14.5

9.3

7.5

5.0

75

19.8

11.7

11.7

8.2

8.7

5.8

100

15.5

10.0

9.9

7.3

9.7

6.5

125

12.9

8.7

8.7

6.6

10.4

7.0

150

11.1

7.8

7.7

6.2

11.1

7.4

175

9.8

7.1

7.1

5.7

11.7

7.8

Note:

Table values are nominal for a 200-joule shock.

T1

T2

I1

I2

I3

I4