LIFEPAK 1000 Defibrillator Operating Instructions

A-3

©2006-2012 Physio-Control, Inc.

Waveform
(continued)

With Infant/Child pads:

The following specifications apply from 25 to 175 ohms.

Energy Accuracy (into 50 ohms):

Selected energy

 4 +/- 15%; 86 joules +/- 15% maximum

Rated energy output is the nominal delivered energy based on the energy
setting and patient impedance, as defined in the following chart.

* Energy setting selected

R a te d E n e rg y O u tp u t

2 0

3 0

4 0

5 0

6 0

7 0

8 0

9 0

2 5

5 0

7 5

1 0 0

1 2 5

1 5 0

1 7 5

P a tie n t Im p e d a n c e ( o h m s )

En

e

rg

y

(J

3 2 5 J *

3 0 0 J *

2 7 5 J *

2 5 0 J *

2 2 5 J *

2 0 0 J *

1 7 5 J *

1 5 0 J *

3 6 0 J *

Rated Energy Output