Pulse Check

The

PULSE CHECK

option inserts prompting to check for a pulse or check the patient,

depending on the

PULSE PROMPT

setting. The choices for

PULSE CHECK

are:

ALWAYS,

AFTER EVERY NSA, AFTER SECOND NSA

, and

NEVER

. The default setting is

NEVER

.

• The

ALWAYS

option prompts for a pulse check after

CPR TIME 1

and

2

, after a

NO SHOCK

ADVISED

decision, after a single

SHOCK ADVISED

decision with

STACKED SHOCKS OFF

, or

after three consecutive

SHOCK ADVISED

decisions if

STACKED SHOCKS

is

ON

.

• The

AFTER EVERY NSA

option prompts for a pulse check after every

NO SHOCK ADVISED

decision.

• The

AFTER SECOND NSA

option prompts for a pulse check after the second analysis if the

second analysis results in a

NO SHOCK ADVISED

decision, regardless of the first analysis

decision (

SHOCK ADVISED

or

NO SHOCK ADVISED

).

• The

NEVER

option eliminates all

PULSE CHECK

prompts.