Changing Setup Options

D-4

LIFEPAK 1000 Defibrillator Operating Instructions

Access Energy Protocols from the AED menu. The underlined bold options in

Table D-5

are the

factory default settings.

Access CPR Setup from the AED menu. The underlined bold options in

Table D-6

are the factory

default settings.

Table D-5

AED Setup Menu—Energy Protocols Submenu

Menu Item

Help Message

Options

ENERGY 1

Select energy level for shock 1.

150, 175,

200

,

225, 250, 275, 300, 325,

360

*

joules

.

*

When selecting 360 joules for energy 1, consider AED use in children.

ENERGY 2

Select energy equal to or
greater than Energy 1.

150, 175, 200, 225, 250, 275, 300

,

325,

360 J

oules

.

ENERGY 3

Select energy equal to or
greater than Energy 2.

150, 175, 200, 225, 250, 275, 300, 325

,

360

joules

.

FLEXIBLE
PROTOCOL

Repeat previous energy after

NO SHOCK ADVISED

(only

when

NO SHOCK ADVISED

follows a shock).

ON

,

OFF

.

STACKED
SHOCKS

Enable consecutive shocks
without CPR.

ON

,

OFF

.

PREVIOUS PAGE

Go back to previous page.

Table D-6

AED Setup Menu—CPR Submenu

Menu Item

Help Message

Options

CPR TIME 1

Set CPR interval after shocks.

15, 30, 45, 60, 90,

120

,

180

seconds

.

CPR TIME 2

Set CPR interval after

NO

SHOCK ADVISED

.

15, 30, 45, 60, 90, 120

,

180

seconds

.

INITIAL CPR

Enable Initial CPR.

OFF

,

ANALYZE FIRST, CPR FIRST

.

INITIAL CPR TIME

Set CPR interval after first
analysis.

15, 30, 45, 60, 90,

120

,

180

seconds.

PRESHOCK CPR

Set CPR interval before

SHOCK ADVISED

decisions.

OFF

,

15, 30

seconds

.

CPR PROMPT

Enable extended CPR prompt:

PROVIDE RESCUE BREATHS
AND CHEST COMPRESSIONS

ON

,

OFF

.

PREVIOUS PAGE

Go back to previous page.