Defibrillation in manual mode, Analysis, Defibrillation in manual mode -6 – Physio-Control LIFEPAK 1000 User Manual

Page 28: Analysis -6

Advertising
background image

3-6

LIFEPAK 1000 Defibrillator Operating Instructions

How to Use the LIFEPAK 1000 Defibrillator

DEFIBRILLATION IN MANUAL MODE

The LIFEPAK 1000 defibrillator provides a Manual mode to override the AED features of the
defibrillator. Manual mode provides operator-initiated analysis, charge, shock, and disarm
functions. This mode is useful in a tiered response system when a provider trained in manual
defibrillation and authorized to place the defibrillator in Manual mode takes over the scene from
a BLS-AED trained provider.

To use Manual mode:

1 Press the Menu button.

2 Select

YES

to enter Manual mode. The ECG trace and Heart Rate Indicator appear on

the screen.

3 If the displayed ECG rhythm appears shockable, press

CHARGE

to initiate charging of

the defibrillator. The screen will indicate that the defibrillator is charging and a charge tone
will sound.

4 Clear everyone away from the patient, bed, or any equipment connected to the patient.

5 When the charge is complete, press the flashing

(shock) button to deliver energy to

the patient.

6 After delivering a shock, the energy for each subsequent shock is automatically selected

based on the energy level configured in Setup.

Note: To remove an unwanted charge at any time, press

DISARM

.

Analysis

The LIFEPAK 1000 defibrillator can be set up to display an

ANALYZE

softkey when in

Manual mode.

To initiate an analysis:

1 Confirm that the patient is unresponsive, not breathing, and without a pulse.

2 Press

ANALYZE

.

3 If the rhythm analysis results in a No Shock Advised decision, the defibrillator remains in

Manual mode without further prompts.

4 If the rhythm analysis results in a Shock Advised decision, the defibrillator automatically

begins charging accompanied by a charge tone. If you determine that a shock is not
warranted, press

DISARM

.

5 When the charge is complete, clear everyone away from the patient, bed, or any equipment

connected to the patient.

6 Press the flashing

(shock) button to deliver energy to the patient.

7 After delivering a shock, the defibrillator remains in Manual mode.