A&D Weighing Indicator AD-4405 User Manual

Page 14

Advertising
background image

Page

12

AD-4405 Weighing Indicator

Note that the

ON/OFF

key does not function alone. Press the

ESC

key while

holding the

ON/OFF

key to end the calibration mode, if mis-operation. After

displaying Can5el, press the

ON/OFF

key to end the calibration mode and

turn the indicator off.

NOTE: When

displaying

Caloff , press the +/- key while pressing the

ON/OFF

key,

instead of

CAL

key. However, in some cases, settings are altered to nullify

this option.

Caution

The maximum display is less than or equal to 40000 divisions. This number

is calculated from the maximum capacity divided by the minimum division.

Check the accuracy of the weighing instrument periodically.

Recommended mass, use a mass heavier than 2/3 maximum capacity.

Calibrate the scale, if it is moved to another location or the environment

has changed.

It is not necessary to set the gravity acceleration correction, when calibrating

the scale with a calibration mass at the place where the scale is used.

Enter the stable weighing data while the STABLE mark is displayed. If

unstable data is used, it may cause a weighing error. Arrange the

condition using the F00 filter function.

The span calibration needs the zero calibration data. We recommend

that you perform the span calibration immediately after the zero

calibration.

If you use the dual range function of the multi-interval scale, perform the

"Range Function", "Zero Calibration" and "Span Calibration".

5.2.

5.2.

5.2.

5.2. Calibration Procedure

Calibration Procedure

Calibration Procedure

Calibration Procedure

5.2.1.

Configuring a Weighing Instrument

This section explains how to set capacity, resolution, decimal point position and display

format, weighing range and unit. Perform this procedure when installing the indicator.

When Cal5et appears on the display, enter the setting mode by pressing of the

ENTER key.

Setting the range and unit.

Single Range

Select the resolution, decimal point position and format.

¯

Specify the weighing capacity.

5. Calibration