A&D Weighing Indicator AD-4405 User Manual

Page 19

Advertising
background image

AD-4405 Weighing Indicator

Page

17

ESC

key ..... The key to store the parameters temporarily. Proceed to the

Cal 0

display.

Press and hold the

ON/OFF

key and press the

ESC

key

No parameters are changed, CanCel is displayed and the calibration

mode is finished.

Step18 Press

the

ON/OFF

key to turn the display off.

5.3.

5.3.

5.3.

5.3. Weighing Range Function

Weighing Range Function

Weighing Range Function

Weighing Range Function

The weighing range function can select "single range" or "dual range". Specify each

weighing interval (division) for the multi-interval instrument. Each weighing interval is

displayed according to a net value or gross value.

Caution

When single range is used, performing this function is not required.

Example 1

The gross display.

Specified

parameters:

First range

Range = 50.00kg, division 0.02kg

Second range

Range = 100.00kg (maximum capacity), division 0.1kg

Display

0kg to 50kg :

The first range, division 0.02kg.

50kg to 100kg :

The second range, division 0.1kg.

Division

=

0.02kg

Division

=

0.1kg

0kg 50kg 100kg Gross

Weighing

value

Example 2

The net display using a 40kg tare value.

Specified parameters:

The same parameters as example 1.

Display

-40kg to 50kg :

The first range, division 0.02kg.

50kg to 60kg :

The second range, division 0.1kg.

Division = 0.02kg

Division=0.1kg

-40kg 0kg 50kg 60kg Net

Tare value

90kg 100kg Gross

40kg Weighing value

5.3. Weighing Range Function

Advertising