Adjustment of the load cell output, Verifying load cell output and input sensitivity – A&D Weighing Indicator AD-4405 User Manual

Page 8

Advertising
background image

Page

6

AD-4405 Weighing Indicator

3.1.3.

Adjustment of the Load Cell Output

Caution

Use a metal film resistor in the range of 50kohm to 500kohm with a

good temperature coefficient, when adding a resistor to adjust a load

cell output. Use as large of a resistance value as possible in the range

in which the zero adjustment is possible. Solder this resistor at a point

near the load cell or the indicator.

In Case of Reducing the Output Voltage

When the zero output is too large, add a resistor

between EXC+ and SIG-.

In Case of Adding an Offset Voltage to the Output

When the zero output is too small, add a resistor

between EXC+ and SIG+.

3.1.4.

Verifying Load Cell Output and Input Sensitivity

The input sensitivity of the indicator is 0.25

mV /division or more. Adapt to the following

inequality, when you design a weighing instrument using the indicator and load

cell(s).

Caution

A change in input voltage sensitivity is required to cause a one division

change of the display. Select as large an input sensitivity voltage as

possible so that the weighing interval becomes stable.

Consider the leverage if a lever is used.

Weighing instrument
using one load cell.

A

D

B

E

*

*

25

.

0

£

Weighing instrument
using multi-load cell

N

A

D

B

E

*

*

*

25

.

0

£

A: Rated capacity of load cell [kg]
B: Rated output [mV/V]
D: Weighing interval [kg]
E: Excitation voltage [mV]
N: Number of load cells

Verification Example

Design:
Load cell

N=1

Rated capacity

A=750 [kg]

Rated

output

B=3

[mV/V]

Excitation voltage E=5000 [mV]
Weighing interval

D=0.05 [kg]

Weighing capacity

300 [kg]

25

.

0

1

750

05

.

0

3

5000

*

*

³

=

. Therefore,

regard the instrument as a good
design.

3. Installation and Precautions

Advertising