Exiting the calibration mode, Page 16 – A&D Weighing Indicator AD-4405 User Manual

Page 18

Advertising
background image

Page

16

AD-4405 Weighing Indicator

Weighing a Mass less than the Maximum Capacity

Step12 Set a mass value using the

0

-

9

keys.

Step13 Place a mass equivalent to the displayed value

on the weighing unit. Proceed to step 15.

Weighing a Maximum Capacity Mass

Step14 Place a mass equivalent to the maximum

capacity on the weighing unit.

Step15 Press the

ENTER

key after the STABLE

mark turns on. Proceed to step17.

ESC

key .......... The key to proceed to step17

without changing the span

parameters.

Caution: Do not press the

ENTER

key while the

STABLE mark is off (detecting motion).

Arrange the condition using the F00 filter

function.


Digital Input
Step16 Pressing

the

+/-

key, a stored input voltage

parameter of the span is displayed in the

unit of mV/V. Adjust the input voltage using

the

0

-

9

keys. (It is possible to store a

greater value than the capacity.)

ENTER

key...... The key to store the span

parameters and proceed to

step 17.

ESC

key .......... The key to proceed to step 17

without changing the span

parameters.

5.2.5.

Exiting the Calibration Mode

Step17 Check

the

Calend display.

Use the following keys.

CAL

key ..... To store the parameters and display

Caloff. Proceed to the next step.

Caloff

could be displayed by pressing

the

+/-

key while pressing the

ON/OFF

key

instead of

CAL

key.

5.2. Calibration Procedure

Turn on STABLE mark

Span calibration

Maximum capacity

To step 14, step 16

To step 12

New mass value

Mass of

maximum capacity

To step 14

To step 16

Place the displayed

mass on the

weighing unit

step 13
step 14

Stored input voltage

New input voltage

mV/V

mV/V

Turn display off

9

0

9

0

+/

+/

+/

+/-

--

-

NET/GROSS

PRINT

ENTER

PRINT

ENTER

PRINT

ENTER

+/

+/

+/

+/-

--

-

NET/GROSS

PRINT

ENTER

or