Page

32

AD-4405 Weighing Indicator

6.3.

6.3.

6.3.

6.3. CF

CF

CF

CF-

-

-

-Functions

Functions

Functions

Functions

Item

Parameter

Description

* 0

No limitation

Cf00
Zero tracking width,
motion detection
condition

1

Use limitation at F01, F02, F03, F27 and F28.

* 0

2% of CAP,

Tare limit is 100 % CAP

1

10% of CAP, Tare limit is 100 % CAP

2

3% of CAP,

Tare limit is 50 % CAP

Cf01
Zero range
Turning display on, the
range to zero display.

3

4% of CAP,

Tare limit is 50 % CAP

0

Not to zero when turning the display on.

* 1

10% of CAP

2

3% of CAP

Cf02
Power on zero range
Turning the display on,
the range to zero display.

3

4% of CAP

0

Gross when displaying gross.

1

Gross

Cf03
Zero tracking

* 2

Gross or Net when displaying net.

TARE, ZERO in motion / TARE at negative gross

* 0

Not accepted

/

Not accepted

1

Execute /

Not

accepted

2

Not accepted

/

Execute

Cf04
TARE, ZERO in motion /
TARE at negative gross
value

3

Execute /

Execute

* 0

Not to output data at unstable value or overload.
Effective in key mode.

Cf05
Output on overload and
unstable state.

1

To output data always.

GROSS / NET / TARE / Preset TARE

* 0

GR / NT / TR /

TR

1

GR / NT / TR /

PT

Cf06
Header 2

2

G_ /

N_ / T_

/

PT (_: Space 20h)

* 0

Two digits

Cf07
Number of unit digits

1

Three digits

Serial output data

* 0

Not used (Ineffective)

Cf08
Accumulation function

1

Used (Effective)

* 0

Used (Effective)

Cf09
Digital Tare(PT)

1

Not used (Ineffective)

CAP : maximum capacity

*: Initial settings

Power on ZERO (CF02)

Cf02 0

After power on, the weight display starts immediately.

Cf02 1,2,3

When the zero range is exceeded at power on, ------ is displayed.

Press

the

ESC

key to start the weight display.

6.3. CF-Functions