A&D Weighing Indicator AD-4405 User Manual

Page 72

Advertising
background image

Page

70

AD-4405 Weighing Indicator

20.

20.

20.

20. Current Loop Output (OP

Current Loop Output (OP

Current Loop Output (OP

Current Loop Output (OP-

-

-

-08)

08)

08)

08)

Replacing the RS-232C interface with this option, current loop output, 3-relay outputs

and 1-control input can be used with the RS-232C interface of this option.

RS-232C functions and pin connections are the same as the RS-232C interface

described in “16. RS-232C Interface”.

The control input can control the indicator from an external terminal just like the front

panel key operations.

Set the external control function at f15 of the F-Functions.

When connecting the function pin and the common pin, the indicator makes the action.

Keep a signal width of more than 100 ms for the On-time and Off-time.

ON(Make)

More than 100 ms More than 100 ms

OFF(Break)

The relays output the result of comparison.

Solid-state-relay

Maximum voltage

DC50V

Maximum current

DC100mA

Maximum resistance 8

9

Pin connections (DIN 8pin connector)

Function

Pin No. Signal name

Direction

Description

1

HI

Output

Relay output HI

6

OK

Output

Relay output OK

4

LO

Output

Relay output LO

Relay

output

8 COM(out) -

Relay

common

terminal

3

CL1

Output

Current loop output 1

Current

loop

5

CL2

Output

Current loop output 2

7

EXT3

Input

Control input 3 (f15)

Control

input

2

COM(in)

-

Control input common

Adaptable connector

JA-TCP0586 (of accessory)

Circuits

Circuits

Circuits

Circuits

Current loop output

HI

OK

LO

COM(out)

K

EXT3

+7 to 9V

680

9

'

COM
(in)

'

DIN connector

of OP-08

Control input Relay output

20. Current Loop Output (OP-08)