Accu-Chek insulin pump User Manual

Page 117

Advertising
background image

Adjusting the Beep Volume

Your pump beeps when a key is pressed or when an alert or error occurs.

Follow the steps below:

1. Press

d to move to the SETUP MENU STANDARD screen.

Press

f to select.

2. Press

d to move to the BEEP VOLUME screen.

The active beep volume is displayed.

3. Press

a or s to adjust the beep volume.

There are five volume levels:

no beeps (beeps are turned off)

very low

low

normal

maximum

4. Press

f to save and exit.

Note:

The volume of the STOP-Warning is independent of the programmed
beep volume. It always occurs at the maximum level. When the beeps
are turned off and an alert or error occurs, the beeps will become active
again after a short period of time to make sure that you are aware of
the alarm or error. These beeps will steadily increase in volume until
the maximum level is reached if the alert or error is not cancelled.

Chapter 5: User Menus

ACCU-CHEK Spirit Insulin Pump

116