Error E10: CARTRIDGE ERROR

The CHANGE THE CARTRIDGE function was not correctly carried out.

1. Press

f twice to turn off and to confirm the error.

2. Remove or disconnect the infusion set from your infusion site.

3. Start the CHANGE THE CARTRIDGE menu.

4. Prime the infusion set.

5. Reconnect the infusion set (tubing) at your infusion site.

6. Put your pump into RUN mode.

7. If an error E10: CARTRIDGE ERROR occurs during the winding of the piston

rod, insert a new battery into the pump and repeat steps 1-6.

Error E11: SET NOT PRIMED

The cartridge was changed but the infusion set was not primed. The error will
display when you try to put your pump in RUN mode.

!

Never prime an infusion set that is connected to your body.
You risk uncontrolled insulin delivery into your body. With
detachable infusion sets, make sure you have disconnected
the tubing from your site prior to priming. Always follow
the instructions provided with your infusion set.

Chapter 7: Alerts and Errors

ACCU-CHEK Spirit Insulin Pump

162

WARNING