Error E2: BATTERY DEPLETED

The battery is out of power. Replace it immediately.

1. Press

f twice to turn off and to confirm the error.

2. Replace the battery.

3. Put your pump in RUN mode.

See the section “Accessories and Disposables–Battery” (page 21) for more
information on recommended batteries. See the section “Inserting (or changing)
the Battery” (page 25) for instructions on how to change your battery.

Note:

Your pump stores the settings you saved before the battery power ran
out. If your pump is without power for more than one hour, check the
time and date.

!

After inserting a new battery, always check that the time
and date of your insulin pump are set correctly. Incorrect
programming of the time and date may cause incorrect
insulin delivery. If you, your doctor, or healthcare team
store and analyze your therapy data electronically and the
time and date of the devices used are not set identically,
the gathered data might not be meaningful.