Accu-Chek insulin pump User Manual

Page 88

Advertising
background image

Make sure that the cancellation was intended and then program a new
Extended Bolus (and/or Temporary Basal Rate), if needed.

Note:

If a Temporary Basal Rate is active at the same time, it is also cancelled
when your pump is put into STOP mode. An alert A8: BOLUS CANCELLED
and an alert A6: TBR CANCELLED occur.

Press

f twice to confirm and turn off the first alert. The second alert

appears on the display. Press

f twice to confirm and turn off the second

alert. Both alerts are recorded in the alarm history.

See the sections “Alert A6: TBR CANCELLED” (page 152) and “Alert A8:
BOLUS CANCELLED” (page 153) for more information.

Chapter 4: Boluses and TBRs

ACCU-CHEK Spirit Insulin Pump

87

Advertising