Automatic Off

Automatic off is a safety feature that stops insulin delivery and triggers an
error E3: AUTOMATIC OFF if no keys are pressed within a programmed time
period in RUN mode.

The automatic off can either be set to OFF or programmed up to 24 hours in
1-hour intervals in the SETUP MENU STANDARD screen.

Follow the steps below:

1. Press

d to move to the SETUP MENU STANDARD screen.

Press

f to select.

2. Press

d to move to the AUTOMATIC OFF screen.

The current status is displayed at the bottom
left of the screen.

3. Press or scroll

a to increase or s to decrease the duration for the

automatic off in 1-hour intervals.

4. If you would like to set the automatic off function to OFF, press or scroll

s

until 0 hours and OFF appears on the display.

5. Press

f to save and exit.

!

Discuss using the automatic off function with your doctor
or healthcare team.