If the error E6: MECHANICAL ERROR occurs again in spite of these actions,

contact your doctor or healthcare team for an alternate therapy plan. Contact

Pump Support at 1-800-688-4578 for more assistance.

Error E7: ELECTRONIC ERROR

An electronic error was discovered during a routine automatic system check
and the insulin delivery has stopped. You cannot turn off and confirm an error
E7: ELECTRONIC ERROR by pressing

f. You must remove the battery instead.

1. Remove or disconnect the infusion set from your infusion site.

2. Insert a new battery into the pump.

3. Prime the infusion set.

4. Reconnect the infusion set (tubing) at your infusion site.

5. Put your pump in RUN mode.

If an error E7: ELECTRONIC ERROR occurs during programming (i.e., basal rate,

time and date) check your settings.

If the error E7: ELECTRONIC ERROR occurs again, contact Pump Support at
1-800-688-4578 for assistance. Contact your doctor or healthcare team if
you need an alternate therapy plan.

Chapter 7: Alerts and Errors

ACCU-CHEK Spirit Insulin Pump

160