Start-up graph over the stabilization period

The start-up graph shows changes in the flow rate over the stabilization time.

The measurements were made at a medium basal rate of 0.1 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.

Bolus
For U100 insulin the delivery accuracy of a maximum bolus is

± 5% and the

delivery accuracy of a minimum bolus is

± 30%.

The measurements were made according to IEC 60601-2-24:1998 with a
Disetronic Classic infusion set PC 16/110 at room temperature.

Delivery Mode

Delivery Rate/Volume

Accuracy

Range

Basal

0.1 U/h

±20%

0.08 - 0.12 U/h

Bolus

0.1 U

±30%

0.07 - 0.13 U