Accu-Chek insulin pump User Manual

Page 168

Advertising
background image

Flow (delivery rate)

0.2 U/sec during infusion set
priming and bolus when using
U100 insulin.

Infusion sets

ACCU-CHEK infusion sets with
luer-lock connector.

IPX 8

Protected against the effects of
temporary immersion in water
under standardized conditions
[up to 60 minutes and 8 feet
(2.5 meters)].

Maximum time for an error

Plastic cartridges*

E4: OCCLUSION

• at med. basal rate 1.0 U/h using

U100 insulin

5h

• at min. basal rate 0.1 U/h using

U100 insulin

50h

Maximum volume* before error

U100 insulin

3.5 U

Maximum pressure

400kPa (4.0 bar)

Maximum quantity delivered at

U100 insulin

1.0 U

a single fault condition

Power supply

One 1.5 volt AA Alkaline battery
or one rechargeable Nickel Metal
Hydride (NiMH) AA battery. Alkaline
batteries should have a minimum
capacity of 2500 mAh and NiMH
AA rechargeable batteries a
minimum capacity of 1500 mAh.
Use only a battery charger
officially recommended by the
battery manufacturer.

Pump casing

Shock, scratch, pharmaceutical
resistant plastic. All edges are
rounded.

*Determined using measurement methods according to IEC 60601-2-24-1998.