Accu-Chek insulin pump User Manual

Page 179

Advertising
background image

Typical Range and

Maximum Parameter

Parameter Settings

Range

Basal Rate
– hourly basal rate +

U100: 62.5 IU/h

Insulin Type

Max. Amount

max. TBR

U100

0 – 62.50 IU/h

(hourly basal rate
combined with temporary
basal rate, max. increase)

– hourly basal rates

0.1 U

increment

– hourly basal rates

U100: 10.0 IU/h

Insulin Type

Max. Amount

U100

0 – 25.0 IU/h

– temporary basal rate

100%

0 – 200%

– TBR increase or decrease

10%

10%

in increments of

– TBR duration

max. 24 h

15 min – 24 h

– TBR duration increments

15 min

15 min, 30 min, 1 h

– basal rate profiles available

5

1 – 5

– basal rate lock

off

on or off

Bolus
– bolus amount per

U100: 25.0 IU/h

Insulin Type

Max. Amount

bolus delivery

U100

0 – 25.0 U

– bolus amount increments

0.5U

0.1, 0.2, 0.5, 1.0 or 2.0 U

for “Quick” and “Scroll”
Standard Bolus

– bolus duration (Extended

max. 24 h

15 min – 24 h

bolus, MultiWave bolus)

– bolus duration increments

15 min

15 min, 30 min, 1h

Appendixes

ACCU-CHEK Spirit Insulin Pump

178