5. Press

f to select. The priming amount of 25 units of insulin is shown on

the display.

6. When START PRIMING is displayed, press and hold

f until you hear a

melody and the piston rod moves forward.

The priming process begins and a count-up is
shown on the display. Your pump stops priming
after 25 units.

When the priming process is finished, your pump
returns to STOP.

The infusion tubing set is properly primed when

there are no air bubbles visible in the tubing and a drop of insulin has come
out of the connector or cannula.

Stopping the Priming Process

You can stop the priming process at any time by
pressing any key. Your pump displays the STOP
PRIMING screen.