Chapter 1
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 10 of 60

The carrier-air inlet pumps extra air to the dosing outlets to speed de-

livery of the tracer-gas to the dosing point. This inlet has a coarse air-
filter, a pump, and a pressure transducer for checking the efficiency of

the pump. Delivering a dose of tracer-gas to a dosing point 50m distant
takes 1 minute.

The dosing system can deliver a continuous dose, that is, an uninter-

rupted flow of tracer-gas over a period of time; or a discontinuous
dose, in which the amount of tracer-gas delivered is reduced by inter-

rupting the flow at regular intervals during the dosing period.

The doser system is set up for use by specifying a dosing time-out such
that the 1303 will stop any current dosing procedure after a given time

has elapsed without instruction from the controlling computer. This val-
ue is specified form the controlling computer.

1.1.3

Front Panel

Fig.1.2 The front panel of the 1303

Doser:

6 mounting stubs for connection of tubing (AF0005, red)
to dosing points. Each stub is numbered, and has a cor-

respondingly-numbered lamp. When the lamp is lit, it
indicates that the corresponding dosing valve is open,

see

section 3.6.3

. When the Main Valve Open lamp is

lit, it indicates that the main dosing valve is open, see

section 3.6.2

.

Sampler:

6 mounting stubs for connection of tubing (AF0006,

green) to sampling points. Each stub is numbered, and
has a correspondingly-numbered lamp. When the lamp