LumaSense Technologies INNOVA 1303 User Manual

Page 16

Advertising
background image

Chapter 2
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 16 of 60

takes the next sample for analysis from the room. As the amount of

tracer-gas delivered to the room is known, and the remaining concen-
tration of tracer-gas in the samples is determined by the Gas Monitor,

the ventilation-system performance can be calculated.

Fig.2.3. A typical air-change analysis system, shown with an applica-

tion example. In this case, the aim of the analysis is to deter-
mine the size of the air-change in the mechanically ventilated

room. The diagram shows only one dosing and sampling point,
for clarity. Up to 6 similar analyses can be performed simulta-

neously using one 1303. 7620 Application Software gives con-
trol of all the functions of the system.

2.2.1

Connecting the 1303 to the System Controller

The 1303 is connected to the computer by the IEEE interface bus, over
which the 1303 receives the commands and data which control it. Fur-

ther information about the cables available from LumaSense to connect
the 1303 onto the IEEE interface bus, and about the IEEE interface of

the 1303, are given in

section 4.1

.

2.2.2

Setting the Interface Address of the 1303

The 1303 uses a single interface address. The address is set using the
bank of eight switches on the rear panel of the 1303. The last five

switches on the bank decide the interface address. These switches are
marked A1 (the least significant bit) to A5 (the most significant bit).

The switches represent binary values, as follows:

If the switch is down, it represents a binary “0”. See

Fig. 2.4

.

Advertising