BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 57 of 60

Table 4.7 Standardised interface jobs which perform various secondary

functions of the 1303.

Job Header

Data

Effect on 1303

Define_Terminator Decimal value of the

ASCII control character

Selects the terminator

character for interface
jobs.

Identify?

No data

Outputs INNOVA 1303

Output_Header

EXclusive
Inclusive

Disables/enables the
1303 to output the mini-

mum code of the job
header with the appro-
priate data, in response

to the output jobs listed,

section 3.2.2

.

*IDN?

No data.

Outputs INNO-
VA,1303,VPXXXX

*RST

No data.

Identical to the job Re-

set_System.

*SRE

As for Ser-
vice_Request_Enable.

Identical to the job Ser-
vice_Request_Enable.

*SRE?

No data

Identical to the job Ser-

vice_Request_Enable?

*STB?

No data

Reads-out the status byte
from the 1303.

*TST?

No data

Indicates if the Warning
or Error flags are set. The
1303’s condition is given

as follows:
0: Normal

1: Warning flag(s) set

-1: Error flag(s) set

The flags can then be
read-out using the Warn-

ing? or Error? jobs.