1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 40 of 60

ponents represented by each bit in the Status flag, and their corre-

sponding decimal values.

Table 3.3. The 1303’s Status Flag. Each bit represents a particular

1303 component; when set, the 1303 is currently using

that component.

Status Flag

Current function when

set

Bit

no.

Dec.

value

Bit name

1

1 Dosing valve 1

Dosing valve 1 open

2

2 Dosing valve 2

Dosing valve 2 open

3

4 Dosing valve 3

Dosing valve 3 open

4

8 Dosing valve 4

Dosing valve 4 open

5

16 Dosing valve 5

Dosing valve 5 open

6

32 Dosing valve 6

Dosing valve 6 open

7

64 Main dosing valve Main dosing valve open

8

128 Dosing pump

Dosing pump on

9

256 Sampling valve 1

Sampling valve 1 open

10

512 Sampling valve 2

Sampling valve 2 open

11

1024 Sampling valve 3

Sampling valve 3 open

12

2048 Sampling valve 4

Sampling valve 4 open

13

4096 Sampling valve 5

Sampling valve 5 open

14

8192 Sampling valve 6

Sampling valve 6 open

15

16384 3-way valve

3-way valve set to ana-

lyzer

16

32768 Sampling pump

Sampling pump on

Example:

STATUS?<LF>

Outputs the Status flag of the 1303

as described above. Refer to

Table

3.3

for the decimal values of each

bit. Refer also to the examples giv-
en below.

If the 1303 outputs 33024, using

Table 3.3

shows that 33024 = 256 +

32768. Therefore, bits 9 and 16 are set. The mechanical condition of
the 1303 is: sampling valve number 1 open, and the sampling pump

running.

If the 1303 outputs 199, using

Table 3.3

shows that 199 = 1 + 2 + 4 +

64 + 128. Therefore, bits 1, 2, 3, 7 and 8 are set. The mechanical con-

dition of the 1303 is: dosing valves 1, 2 and 3 open; main dosing valve
open; dosing pump running.