3 the calibration data parameter – LumaSense Technologies INNOVA 1303 User Manual

Page 31

Advertising
background image

Chapter 3
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 31 of 60

Alternatively, the gas constant can be expressed in terms of the molec-

ular weight of the gas:

MOL_WEIGHT 146,05<LF> Sends the gram molecular weight

146,05 (for SF

6

) to the 1303.

The current characteristic gas constant or molecular weight value (de-

pending on how you have previously expressed the gas constant to the
1303) can be checked using the following jobs:

GAS_CONSTANT?<LF>

Outputs the current gas constant

value from the 1303.

MOL_WEIGHT?<LF>

Outputs the current molecular

weight value from the 1303 in kg
kmol

-1

(equivalent to g mol

-1

).

Table 3.2. Molecular weight of commonly used Tracer-gases

Tracer Gas

M

R

/

M

Molecular Weight

g/mol = kg/kmol

Characteristic Gas

Constant

SF

6

146.05

56.92

Freon 134a

102.03

81.49

Freon 152

66.05

125.87

CO

2

44.01

188.91

N

2

O

44.01

188.91

3.3.3

The Calibration Data Parameter

This specifies the effective outflow area (in m

2

x 10

-9

) of a specific dos-

ing nozzle, thus allowing the 1303 to automatically calculate the

amount of tracer-gas delivered during a dosing procedure. The 1303 is
able to calculate these values automatically for each dosing nozzle

when calibrating itself – see

section 3.4

for further information. A dos-

ing nozzle cannot be used until it is calibrated with the tracer-gas to be

used.

The range of values is from 0,1 to 100,0; there is no default value for
this parameter.

If the calibration data for each specific nozzle and for the tracer-gas

currently being used is known, it can be sent directly to the 1303 using
the following job, without calibrating the 1303. Note that the 1303’s

doser nozzles must still be calibrated with each change of tracer-gas,
so that the 1303 is able to calculate the amount delivered.

Advertising