4.2.1

Set-up Jobs

These jobs install set-up parameters in the 1303. The set-up parame-

ters must be installed before starting to use the 1303. Further informa-
tion about the set-up parameters is given in

section 3.3

.

Table 4.1 The interface jobs which set-up the 1303.

Job Header

Data

Effect on 1303

Dosing_Time_Out

Time interval form

10s to 3600s.

Sends the new value for

the Dosing Time-out pa-
rameter to the 1303.

Dosing_Time_Out? No data.

Reads-out the current

Dosing Time-out value
from the 1303.

Gas_Constant

Ratio of R (8314 J

mol

-1

K

-1

) and the

molecular weight M

of the tracer-gas be-
ing used.

Sends the new value for

the Characteristic Gas
Constant parameter to

the 1303.

Gas_Constant?

No data.

Reads-out the current

Characteristic Gas Con-
stant parameter from

the 1303.

Mol_Weight

Molecular weight of
the tracer-gas being

used.

Sends the new value for
the molecular weight of

the tracer-gas being
used to the 1303.

Mol_Weight?

No data.

Reads-out the molecular

weight of the tracer-gas
being used from the

1303.

Calibration_Data

Doser-nozzle number
and area of nozzle in

m

2

* 10

-9

.

Sends the calibration da-
ta for the stated doser

nozzle to the 1303.

Calibration_Data?

Doser-nozzle num-

ber.

Reads-out the calibra-

tion data for the stated
doser nozzle from the
1303.