2 system use – LumaSense Technologies INNOVA 1303 User Manual

Page 15

Advertising
background image

Chapter 2
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 15 of 60

Fig.2.2. Inserting the fuses into the 1303’s fuse holder

Use two 1,6A slow-blow fuses (LumaSense No. VF0007); Fuses are

supplied with the instrument.

When replacing the fuse-holders, ensure that the direction of the white
arrows on each holder matches the arrows marked on the covering

flap.

2.2 System Use

The 1303 combines with the Gas Monitor and a controlling computer to
provide a system which offers wide-ranging monitoring capabilities. The

1303 makes it possible to perform multi-point air-exchange analyses
and multi-point monitoring tasks in many different situations and envi-

ronments, without changing the system components.

An example air-exchange analysis system is shown in

Fig. 2.3

. In such

a system, the doser/sampler systems of the 1303 are used as follows.
The doser system marks the supply-air of the room with a known

amount of tracer-gas. The sampler system then takes a sample of the
return-air from the room, and delivers the sample to the Gas Monitor

for analysis. While the Gas Monitor performs one analysis, the 1303