LumaSense Technologies INNOVA 1303 User Manual

Page 33

Advertising
background image

Chapter 3
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 33 of 60

1303’s temperature to stabilize in relation to the temperature of

the area in which it will be used.

The main dosing valve and each dosing valve must be opened to
allow tracer-gas to flow into the 1303’s dosing manifold, for 5

minutes before calibrating. This ensures that the temperature of
the dosing system will stabilise relative to the temperature of the

tracer-gas. The procedures for opening the main dosing valve and
the dosing valves, to allow flow of tracer-gas through the dosing

system, are described in

section 3.6.2

and

3.6.3

respectively. To

ensure that the tracer-gas does not flow into the room in which you

will be performing measurements, direct the outlets of the doser
tubing to a suitable exhaust – for example, out of a window. After

flushing is complete, wait for 15 minutes before proceeding with
the calibration.

2. If you are to use a different tracer-gas from the one used previ-

ously, you must first flush the doser system with the new tracer-

gas, before calibrating with the new tracer-gas. Flushing is done as

described above. Once flushing is completed, remember to change

the characteristic gas-constant value stored in the 1303 to that for

the new tracer-gas before performing the calibration. This proce-

dure is described in

section 3.3.2

.

Examples:

CALIBRATE_NOZZLE 1<LF> Starts the 1303’s self-calibrating

procedure for dosing nozzle number

1. Each nozzle calibration takes ap-
proximately 5 minutes. If no dosing

nozzle number is given, the 1303
will calibrate all the dosing nozzles.

Each of the dosing nozzles that you wish to use must be calibrated in

this way. Once the procedure is complete, the new calibration data for
each nozzle can be read-out using the following job:

CALIBRATION_DATA? 1<LF> Outputs the calibration data for dos-

ing nozzle number 1. If no nozzle
number is given, the calibration da-

ta for all 6 nozzles is output.

Note: the 1303 does not maintain the calibration data in its memory
after switching-off. It is recommended that you read-out and note

down the current calibration data for each nozzle, for future reference
in the event of loss of calibration data from the 1303. The calibration

data can then be re-installed using the Calibration_Data job,

section

3.3.3

.

Advertising