Chapter 3
________________________________________________________________________

_______________________________________________________________________

BE1085-14

1303 Multipoint Sampler and Doser

LumaSense Technologies A/S

Page 36 of 60

MAIN_DOS_VALVE OP<LF> Opens the main dosing valve to al-

low the flow of tracer-gas to the
dosing valves.

The main dosing valve will close automatically to protect the doser-

system airways if the tracer-gas pressure in the dosing manifold is
above 550 kPa absolute.

3.6.3

Starting Continuous Dosing

Using the interface job described below starts an uninterrupted flow of
tracer-gas through the specified dosing valves. Any number of the dos-
ing valves can be opened simultaneously. The valve specified in the in-

terface job will be opened; those not specified are closed. The 1303’s
doser system is also equipped with a pump which pumps carrier-air

through the dosing valves to accelerate delivery of the tracer-gas to the
dosing points. The dosing pump is controlled separately from the dos-

ing valves.

Note: the carrier-air inlet on the rear panel of the 1303 must not be
blocked.

Example:

OPEN_DOSING_VALVE

1,2<LF>

Opens only dosing valves 1 and

2. If no dosing valves are speci-
fied, all dosing valves are closed.

The dosing pump must then be started. The pump is controlled using
the following commands.

DOSING_PUMP ON<LF>

Starts the carrier-air pump.

DOSING_PUMP OFF<LF>

Stops the carrier-air pump.

If no further instructions are sent to the 1303, the dosing procedure

will be stopped automatically when the time-period specified by the
Dosing Time-out parameter,

section 3.3.1

, has elapsed. To continue the

task in progress, the same job should be sent again to the 1303 before
the dosing time-out period has elapsed.

The carrier-air pump can be used to effect air-change in the casing of

the 1303, to prevent a build-up of the concentration of tracer-gas in
the casing of the 1303. This is done using the interface job DOS-

ING_PUMP AUTO. When sent to the 1303, this job causes the doser
pump to start automatically whenever the pressure in the 1303’s doser