3.8 Checking the 1303

These interface jobs are for checking the various functions of the hard-

ware and software components of the 1303, and help to identify faults
in the 1303.

3.8.1

Sampling_Pump_Pressure? Interface Job

This job reads-out the pressure-difference across the sampling pump.

The pressure across the pump can be measured with the sampling
valves either open or closed. The internal 3-way valve must be set to

Waste Air Outlet.

Example:

SAMP_PUMP_PRESS?<LF> Measures and outputs the pressure-

difference across the sampling

pump, in kPa. With all sampling
valves closed, the pump should

generate a pressure-difference
greater than 40kPa. With the sam-

pling valves open, the pressure-
difference should be less than 25

kPa.

3.8.2

Dosing_Pump_Pressure? Interface Job

This job reads-out the pressure-difference across the dosing pump,
which supplies the carrier-air for dosing.

Example:

DOS_PUMP_PRESS?<LF>

Measures and outputs the pressure-

difference across the dosing pump,
in kPa. This value should be at least

10kPa.

3.8.3

Status? Interface Job

This job reads-out the 1303’s current mechanical condition. The 1303’s
current condition is shown as a “flag”, i.e. a word-value comprising 16

bits, each of which represents a specific mechanical component. If a bit
is set, the 1303 is currently using the corresponding component. The

flag is output from the 1303 as the total of the decimal equivalents of
the binary values of the bits which are set. See

Table 3.3

for the com-