6. Adjusting waveform positions

Selecting the

menu

Press the OSC OPTION key 
repeatedly until the following 
menu appears on the display. 

 – Time 
 – CH1 (or CH2) Zero 

OSC

OPTION

Vertical

center

position

Horizontal

center

position

Use the Left/Right key to change the horizontal 
position. 

Adjusting the

horizontal

position

Original

Use the Up/Down key to change the vertical 
position. 

Adjusting the

vertical position

Original

GDS-122 User Manual

22

5B

Configurations

Overview

The configuration chapter describes how to 
change various GDS‐122 internal parameters for 
allowing better measurement condition. 

Configuration

items

Channel (vertical) settings 

Horizontal settings 

Trigger settings 

Acquisition modes 

Display settings 

System status (only for viewing)

page

120H

22 

page

121H

25 

page

122H

25 

page

123H

30 

page

124H

36 

page

125H

37 

25B

Configuring channel (vertical) settings

Overview

The channel settings configure how the waveform 
appears in terms of vertical or voltage scale. 

Position 

Sets the vertical position of the 
waveform. 

Scale 

Sets the vertical scale (volts per 
graticule). Range: 5mV/div to 5V/div 

CH on/off  Turns the channel on or off. 

Coupling  Selects AC or DC coupling. The DC 

coupling shows all signal elements, 
while the AC coupling filters out the 
DC component from the waveform. 

Inversion  Flips the waveform upside down. 

Magnifica
‐tion 

Magnifies the displayed units (does not 
magnify the real signal). The 
magnification function is useful to 
align the displayed with probe 
attenuation (page

126H

17), especially X10. 