Measuring current, 208h, 209h – GW Instek GDS-122 User Manual User Manual

Page 44

Advertising
background image

Using

the

Multimeter

87

14B

Measuring Current

DC current

specifications

(details: page

208H

98)

Range &

Accuracy


Max input

40mA ±(1% + 1 digit) 
400mA ±(1.5% + 1 digit) 
20A ±(3% + 3 digits) 
400mA (direct input) 
20A (via the extension module) 

AC current

specifications

(details: page

209H

98)

Range &

Accuracy


Max input

40mA ±(1.5% + 3 digits) 
400mA ±(2% + 1 digit) 
20A ±(5% + 3 digits) 
400mA (direct input) 
20A (via the extention module) 

Panel operations

1. Press the A switch to select the 

current (Ampere) measurement. If 
a warning message appears, press 
any key to resume measurement. 

A
A

Battery level

Measurement

unit
Scaling factor

F1-F5 menu

Auto/Manual

range

Run/Stop
Delta

measurement

result

Measurement

result

Meter

-83.60mA

DCA/ACA

2. Press the AUTOSET key 

repeatedly to select DC or 
AC current measurement. 

AUTOSET

Measuring 0mA

to 400mA

1. Connect the test leads to the 

terminals: 
COM for the black lead 
mA/A for the red lead 

400mA

MAX

mA/A

COM

(black)

(red)

If the range is set at 20A, 
press F4 (mA) and change it 
to mA range. 

F2

mA

20A

mA

GDS-122 User Manual

88

2. Select automatic range by 

pressing F3 (Auto) or manual 
by F1 (Manual). The indicator 
at the top left corner of the 
display changes accordingly. 

F1

Manual

F3

Auto

MANUAL

AUTO

Measuring

400mA to 20A

1. Connect the Current 

Extension module to the 
COM & mA/A terminals. 
Then, connect the test leads 
to the extension module. 

20 AMP

400mA

MAX

mA/A

COM

(black)

(red)

2. Press F5 (20A) and select the 

20A range. The MANUAL 
range indicator activates. 
(Auto range is not available) 

F5

20A

MANUAL

Freezing the

measurement

To freeze the measurement, press the 
RUN/STOP key. The measurement 
result will be retained and the 
indicator at the top right corner of the 
display changes to STOP. To unfreeze, 
press the RUN/STOP key again. 

RUN/STOP

RUN

STOP

Measuring delta

current

To measure the delta value, press F2 
(II/

Δ). The measurement result at the 

moment moves to the top right corner 
of the display, and the measurement 
result becomes the difference between 
the original result. 

2. Measurement

reset to zero

20.0mA

1. Press F2

3. Delta value

displayed