Using

the

Oscilloscope

47

38B

Viewing waveforms in X-Y format

Overview

The X‐Y format plots the CH1 input as X‐axis and 
CH2 input as Y‐axis. This display mode is 
convenient for viewing the phase relationship 
between CH1 and CH2. 

Panel operations

1. Make sure that both CH1 and CH2 waveforms 

appear in the display. 

2. Press the MENU key and 

select the DISP SET menu 
using the Up/Down keys. 

MENU

DISP SET

3. Press F3 (Format) and select 

XY. The display mode 
switches into the X‐Y format. 

XY

Format

F3

Changing the

scale and

position

Press the OSC OPTION key 
repeatedly to access the menu listed 
below. In the X‐Y mode, all scales and 
positions are controlled by the 
Up/Down keys. 

OSC

OPTION

CH1 Zero: horizontal position 

CH2 Zero: vertical position 

CH1 Vol: horizontal scale 

CH2 Vol: vertical scale 

Functions not

applicable in the

X-Y format

The following functions do not work in the X‐Y 
format. 

Cursor measurement (page

168H

42, page

169H

43) 

Automatic measurement (page

170H

41) 

Window zoom (page

171H

45) 

Trigger settings configuration (page

172H

25) 

GDS-122 User Manual

48

39B

Viewing signal peaks

Overview

Using the peak detect acquisition mode, the 
maximum and minimum data in the sampling 
interval are displayed, capturing the rapid 
changes and sudden peaks that might 
spontaneously occur in a waveform. 

Note

Since the peak detect mode picks up the most 
extreme data, the waveform becomes noisier than 
the normal acquisition mode (sampling mode). 

Panel operations

1. Press the MENU key and use the Up/Down 

keys to select ACQU MODE menu. 

2. Press F2 (Peak Detect) to activate the peak 

detect mode. 

MENU

ACQU MODE

Peak

Detect

F2

For other acquisition settings details, see page

173H

30. 

Example

Peak detect off

Peak detect on