Using

the

Oscilloscope

31

 

Example:

falling

edge

Triggering point

Falling edge

Selecting the

source channel

Press F3 (Channel) repeatedly to select 
the trigger source channel, CH1 or 
CH2. The trigger channel indicator at 
the bottom right corner of the display 
changes. 

Trigger source

channel

CH1

Source

CH2

F3

Selecting the

trigger mode

Press F4 (Mode) repeatedly to select 
the trigger mode. The trigger status 
icon in the upper right corner of the 
display changes accordingly. For the 
overview of trigger status in general, 
see page

135H

27. 

Single

Mode

Auto

Normal

F4

Auto mode

In the auto mode, input signals are constantly 
acquired and shown in the display regardless of 
trigger condition. 

100ms

5s

5ns

AUTO

Trigd

Scan

Status icon when triggered

Status icon when not triggered

Scan

Horizontal scale (/div)

GDS-122 User Manual

32

Normal mode

In the normal mode, input signals are shown in 
the display only if the trigger condition is met. 

100ms

5s

5ns

Trigd

Scan

Status icon when triggered

Status icon when not triggered

Scan

Horizontal scale (/div)

Ready

Single mode

In the single mode, you manually trigger the GDS‐
122 by pressing the RUN/STOP key each time you 
need to observe the waveform. Once the waveform 
is captured, the GDS‐122 stops triggering and 
waits for the next trigger command. 

STOP

100ms

5s

5ns

Status icon when triggered

Status icon when not triggered

Scan

Horizontal scale (/div)

STOP

AUTO

Ready

Selecting the

coupling mode

Press F5 (Coupling) repeatedly to select 
the trigger coupling. 

AC: triggers only on the AC portion 
of the waveform. 

DC: triggers on the whole waveform 
(AC + DC). 

LF Rjc: filters out the lower 
frequency when triggering. 

HF Rjc: filters out the higher 
frequency when triggering. 

DC

Coupling

AC

LF Rjc

HF Rjc

F5