Measuring Capacitance

Continuity

specifications

Range

51.2nF to 100uF ±(3% + 3 digits) 

Panel operations

1. Press the R switch. If a warning 

message appears, press any key to 
resume measurement. 

2. Press the AUTOSET key repeatedly 

to select the capacitance 
measurement. 

AUTOSET

3. The capacitance screen appears. 

Battery level

Measurement

unit
Scaling factor

F1-F3 menu

(F1, F3 not

used)

Auto range

Run/Stop
Delta

measurement

result

Measurement

result

Meter

Capacitance

Measuring 5nF

and above

Connect the test leads to the CX 
terminals. 

CX

CATII

400V

Measuring less

then 5nF

Connect the Capacitance 
Extension module to the COM 
& V/

Ω/C terminals. Then, 

connect the test leads to the 
extension module. The 
measurement result will be 
constantly updated in the 
display. 

CX <5nf

400V

400V

MAX

COM

V/O/C

GDS-122 User Manual

94

 

Freezing the

measurement

To freeze the measurement, press the 
RUN/STOP key. The measurement 
result will be retained and the 
indicator at the top right corner of the 
display changes to STOP. To unfreeze, 
press the RUN/STOP key again. 

RUN/STOP

RUN

STOP

Measuring delta

capacitance

To measure the delta value, press F2 
(II/

Δ). The measurement result at the 

moment moves to the top right corner 
of the display, and the measurement 
result becomes the difference between 
the original result. 

2. Measurement

reset to zero

59.6nF

1. Press F2

3. Delta value

displayed