Calibration

Overview

The self calibration automatically adjusts the GDS‐
122 internal parameters. The probe calibration 
adjusts the probe capacitance. You should run 
both whenever using the GDS‐122 in a new 
environment. 

42B

Running the self calibration

Overview

The self calibration function automatically 
configures internal parameters to maintain the 
sensitivity and accuracy. Run the self calibration in 
the following cases. 

When the temperature fluctuates more than 5 
degrees Celsius during operations 

When operating the GDS‐122 in a new 
benchtop or field environment 

Procedure

1. Press the MENU key and 

select the FUNCTION menu 
using the Up/Down keys. 

MENU

FUNCTION

2. Press F2 (Autocalibration). A 

message appears on the 
display, asking you to remove 
all cables and probes from the 
GDS‐122. 

Autoca-

libration

F2

3. After removing all cables, 

press F2 (Autocalibration) 
again. The self‐calibration 
automatically starts and a 
message appears, showing 
that the calibration is 
ongoing. 

Autoca-

libration

F2

4. When the message disappears in 5 minutes, the 

calibration is completed. 

To interrupt

calibration

Press any key during the calibration. 

GDS-122 User Manual

52

43B

Running the probe calibration

Overview

The attached probe contains a calibration point at 
the end to adjust the waveform. 

Procedure

1. You may use the GDS‐122’s own 

1kHz square wave output 
signal. Insert the signal cable 
(included in the package) to the 
output terminal. 

2. Use the Autoset function to put 

the waveform in the middle of 
the display. 

AUTOSET

3. Adjust the probe calibration point to make sure 

that the waveform edge remains flat. 

Over-

compensation

Optimum

Under-

compensation