Pecifications – GW Instek GDS-122 User Manual User Manual

Page 49

Advertising
background image

Specifications

97

S

PECIFICATIONS

Conditions for

specifications

The following specifications are applicable when 
these two conditions are met: 

The GDS‐122 has been powered up for at least 
30 minutes, during which temperature 
fluctuation is no more than 5 degrees Celsius. 

The probe attenuation is set to X 10. 

69B

Oscilloscope specifications

Sampling

Mode

Normal, Peak detection, Average

Rate

100 MSa/s

Input Coupling DC,

AC

Impedance

1MΩ±2% in parallel with 20pF±3pF

Probe

1X, 10X, 100X, 1000X

Max. Input

400V (peak)

Channel delay

150ps (typical)

Horizontal

Sampling rate

10S/s~100mS/s

Interpolation

sin x)/x

Record length

6K points on each channel

Scanning

speed

5ns/div~5s/div, 1-2.5-5 step

Sampling rate /
relay time accuracy

±

100ppm (time interval ≥ 1ms)

Interval (△T)
accuracy (full
bandwidth)

Single: ±(1 interval time +100ppm×
reading+0.6ns) Average >16 : ±(1
interval time +100ppm×reading+0.4ns)

Vertical

A/D converter

8 bits resolution (2CH simultaneously)

Sensitivity

5mV/div~5V/div(at input)

Displacement

±

50V(500mV~5V),±1V(5mV~200mV)

Bandwidth

20M

Single

Full bandwidth

Low

frequency

≥5Hz(at input, AD coupling, -3dB)

Rise time

≤17.5ns (at input, typical)

DC

accuracy

±

5% (DC gain)

DC accuracy (avg) Avg >16: ±(5% rdg + 0.05 div) for △V

GDS-122 User Manual

98

Trigger Sensitivity CH1 and CH2: 1div(DC~full bandwidth)

DC coupling: ≥ 50Hz.

Trigger level

±

6 divisions from the screen center

Level accuracy

±

0.3 div (typical, rise/fall time ≥ 20ns)

Displacement

655div (pre-trigger), 4div (post- trigger)

50% level setting Input signal frequency ≥ 50Hz (typical)

Trigger sensitivity 2 div of peak-to-peak (video trigger)

Signal system

NTSC, PAL, SECAM (any frequency)

Measurement Cursor

V and △T between cursors

Automatic

Peak-to-peak, average, root mean
square, frequency, and cycle

Probe

1X position

10X position

Bandwidth

≤ 6 MHz (DC)

Full bandwidth (DC)

Attenuation rate

1: 1

10: 1

Compensation

10pf~35pf 10pf~35pf

Input impedance 1MΩ±2%

10MΩ±2%

Input

impendence

85pf~115pf 14.5pf~17.5pf

Input voltage

150 V DC

300V DC

70B

Multimeter specifications

VDC Input

impedance

10MΩ

Max input

1000V (DC or AC peak-to-peak value)

Accuracy

±

1%±1 digit

Resolution

400mV range: 100uV
4V range: 1mV
40V range: 10mV
400V range: 100mV

VAC Input

impedance

10MΩ

Max input

750V(AC, virtual value)

Frequency

range

40Hz~400Hz

Display

Virtual value of sine wave

Accuracy

±

1%±3 digits

Resolution

4V range: 1mV
40V range: 10mV
400V range: 100mV

DCA Accuracy 40mA range: ±1%±1 digit

400mA range: ±1.5%±1 digit
20A range: ±3%±3 digits

Resolution

40mA range: 10uA

Advertising