Jenway 6850 User Manual

Page 17

Advertising
background image

17

h) Flow-through cuvettes must be selected with care and consideration for the sample

type, sample volume, pumping system, rinse, sample and waste handling to be used.

5. Samples and standards should not be stored in open cuvettes or sample containers as

evaporation will change the value and lead to staining of the walls which may be

irreversible. If stored in stoppered and sealed cuvettes, they should be filled with little or

no air space and the values regularly checked against a reference standard or quality

control material.

6. Samples should be allowed to equilibrate to ambient temperature before measurement

(unless a suitable temperature controlled sample holder is in use). Temperature change

during measurement may cause air bubbles to form on the walls of the sample holder.

This is a common cause of drift during measurement.

7. In the preparation of samples and standards high grade borosilicate glass and AR grade

chemicals and reagents must be used. Good quality deionised water or other suitable

solvents must be used for dissolving or diluting samples, chemicals and reagents.

8. All measurements require calibration to a blank, for maximum accuracy this should be

prepared with care using the same deionised water or solvent used for dissolving or

diluting the sample. Where reagents are added to the sample to produce a colour

propor

tional to its concentration a ‘sample based’ blank should be used. In this case the

blank should consist of all reagents or chemicals to be used, except the sample which

will produce the colour to be measured.

9. Deviations from the Beer-Lambert Law may occur at high and low concentrations giving

non-linear response during sample concentration measurements. For all new methods a

linear range should be defined by the preparation of a calibration curve. The quantitation

mode may be used to construct such a curve against which sample results are

automatically measured.

10. Cuvettes and sample holders must be filled to a minimum level which covers the light

path. All Jenway spectrophotometers have a beam height of 15mm.

11. The instrument must be calibrated to zero absorbance/100% transmittance prior to taking

readings. In the spectrum measurement mode a baseline scan must be performed before

performing a sample scan.

Advertising